Opened 17 months ago
Last modified 13 months ago
#111 accepted requirement
Wiederholbarkeit
| Reported by: | (none) | Owned by: | Erik Welander |
|---|---|---|---|
| Priority: | major | Component: | Gesamtsystem |
| Keywords: | Cc: | ||
| Data type: | Deadline: | ||
| Depending on: | 181 | Implements: | |
| Memaddress: | Observerlist: | ||
| Parent: | #49 | Answer: | |
| Type: | system | Revieweraccept: | welander |
| Reviewerlist: | welander,rotaru | Reviewerreject: | |
| Priority: | normal | Visibility: | |
| Writable: | Class: | quality | |
| Priority: | must have |
Description (last modified by )
Das Gerät mit 6 Pfaden, Class 1.0, soll mindestens eine Wiederholbarkeit haben nach folgendem Schema:
0.1: Bei Hochdruck, zwischen QT & QM
0.3: Bei Hochdruck, unterhalb QT
0.2: Bei atmosphärischem Druck, oberhalb QT
0.5: Bei atmosphärischem Druck, unterhalb QT
Für Class 0.5 muss die Wiederholbarkeit folgendes Schema einhalten:
0.17: Zwischen QT & QM
0.33: Unterhalb QT
unabhängig von Betriebsdruck.
Kommentar:
Werte sind Absolut (i.e. nicht ± 0.1 %).
Dies entspricht die Daten für GT400. Anforderung nach ISO 17089 sind damit auch abgedeckt (0,33 bzw. 0,66) und auch AGA9 (±0,2 % bzw. ±0,4 % entsprechen 0.4 bzw. 0.8).
Bei anderen Pfadauslegungen müssen die Grenze unter Umständen angepasst werden.
Kommentar aus Workshop:
"Class 0.5 Wiederholbarkeit 0.05 vielleicht nur möglich mit 8 Pfade
3 Pfade: Günstiger, Große Nennweiten"
Change History (14)
comment:1 by , 17 months ago
| Description: | modified (diff) |
|---|---|
| Reporter: | removed |
comment:2 by , 17 months ago
| Description: | modified (diff) |
|---|
comment:3 by , 17 months ago
| Description: | modified (diff) |
|---|---|
| Reviewerlist: | welander → welander,rotaru |
| Status: | new → submitted |
comment:4 by , 17 months ago
| Revieweraccept: | → welander |
|---|
comment:5 by , 17 months ago
| Status: | submitted → new |
|---|
comment:6 by , 17 months ago
| Type: | market → system |
|---|
comment:7 by , 15 months ago
| Status: | new → submitted |
|---|
comment:8 by , 15 months ago
| Status: | submitted → accepted |
|---|
comment:9 by , 14 months ago
| Status: | accepted → modified |
|---|
comment:10 by , 14 months ago
| Description: | modified (diff) |
|---|---|
| Status: | modified → resubmitted |
comment:11 by , 14 months ago
| Parent: | #9 → #49 |
|---|
comment:12 by , 14 months ago
| Status: | resubmitted → accepted |
|---|
comment:13 by , 14 months ago
| Depending on: | → 185,191 |
|---|
comment:14 by , 13 months ago
| Depending on: | 185,191 → 181 |
|---|
![(please configure the [header_logo] section in trac.ini)](/rsm600_req.bak/chrome/site/rmglogo.png)