Timeline



Aug 29, 2024:

11:58 AM Ticket #13 (Genauigkeit nach OIML R 137-1 & 2 (2014)) updated by Erik Welander
Revieweraccept changed
11:58 AM Ticket #25 (Elektromagnetische Schutzklasse) updated by Erik Welander
Revieweraccept changed
11:58 AM Ticket #26 (Zugriff Lokal) updated by Erik Welander
Revieweraccept changed
11:37 AM Ticket #112 (Drahtlose Kommunikation) updated by Erik Welander
Revieweraccept changed
11:37 AM Ticket #112 (Drahtlose Kommunikation) updated by Erik Welander
Description, Reviewerlist, Status changed
11:26 AM Ticket #111 (Wiederholbarkeit) updated by Erik Welander
Revieweraccept changed
11:26 AM Ticket #111 (Wiederholbarkeit) updated by Erik Welander
Description, Reviewerlist, Status changed
10:32 AM Ticket #110 (Anschluss für Temperatursensor) updated by Erik Welander
Type, Revieweraccept changed
Wird als Systemanforderung eingestuft.
10:31 AM Ticket #110 (Anschluss für Temperatursensor) updated by Erik Welander
Description, Reviewerlist, Status changed
Austausch mit Jörn am 29.08.2024 wegen Eindringtiefe und …
10:14 AM Ticket #107 (Umgebungstemperatur) updated by Erik Welander
Revieweraccept changed
10:14 AM Ticket #107 (Umgebungstemperatur) updated by Erik Welander
Description, Reviewerlist, Status changed
10:07 AM Ticket #22 (Erweiterte Umgebungstemperatur) updated by Erik Welander
Revieweraccept changed
10:06 AM Ticket #22 (Erweiterte Umgebungstemperatur) updated by Erik Welander
Description, Reviewerlist, Status changed
10:03 AM Ticket #96 (Anschlüsse für Drucksensoren) updated by Erik Welander
Revieweraccept changed
10:03 AM Ticket #96 (Anschlüsse für Drucksensoren) updated by Erik Welander
Reviewerlist, Reporter, Status changed
9:55 AM Ticket #76 (Zertifizierung und Normen) updated by Erik Welander
Status changed
9:55 AM Ticket #76 (Zertifizierung und Normen) updated by Erik Welander
Reporter, Status changed
9:55 AM Ticket #75 (Open Platform Communication (OPC)) updated by Erik Welander
Revieweraccept changed
9:55 AM Ticket #75 (Open Platform Communication (OPC)) updated by Erik Welander
Reviewerlist, Reporter, Status changed
9:54 AM Ticket #65 (Markt) updated by Erik Welander
Revieweraccept changed
9:54 AM Ticket #65 (Markt) updated by Erik Welander
Reviewerlist, Status changed
9:53 AM Ticket #64 (Cry-out) updated by Erik Welander
Revieweraccept changed
9:53 AM Ticket #64 (Cry-out) updated by Erik Welander
Description, Reviewerlist, Reporter, Status changed
9:51 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Revieweraccept changed
9:51 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Reviewerlist, Status changed
9:50 AM Ticket #43 (Standardnennweiten) updated by Erik Welander
Revieweraccept changed
9:50 AM Ticket #43 (Standardnennweiten) updated by Erik Welander
Description, Reviewerlist, Status changed
9:49 AM Ticket #42 (Einlauf- und Auslaufstrecke) updated by Erik Welander
Revieweraccept changed
9:49 AM Ticket #42 (Einlauf- und Auslaufstrecke) updated by Erik Welander
Reviewerlist, Reporter, Status changed
9:48 AM Ticket #41 (Bauform) updated by Erik Welander
Status changed
9:48 AM Ticket #41 (Bauform) updated by Erik Welander
Reporter, Status changed
9:48 AM Ticket #30 (Umwerter über Encoderschnittstelle anschließen) updated by Erik Welander
Revieweraccept changed
9:48 AM Ticket #30 (Umwerter über Encoderschnittstelle anschließen) updated by Erik Welander
Reviewerlist, Reporter, Status changed
9:40 AM Ticket #29 (Umwerter über Hochfrequenzausgang anschließen) updated by Erik Welander
Revieweraccept changed
9:40 AM Ticket #29 (Umwerter über Hochfrequenzausgang anschließen) updated by Erik Welander
Reviewerlist, Status changed
9:39 AM Ticket #28 (Umwerter über Niederfrequenzausgang anschließen) updated by Erik Welander
Revieweraccept changed
9:39 AM Ticket #28 (Umwerter über Niederfrequenzausgang anschließen) updated by Erik Welander
Reviewerlist, Reporter, Status changed
9:39 AM Ticket #27 (Umwerter lokal anschließen) updated by Erik Welander
Status changed
9:39 AM Ticket #27 (Umwerter lokal anschließen) updated by Erik Welander
Reporter, Status changed
9:37 AM Ticket #26 (Zugriff Lokal) updated by Erik Welander
Description, Reviewerlist, Status changed
9:36 AM Ticket #25 (Elektromagnetische Schutzklasse) updated by Erik Welander
Reviewerlist, Status changed
9:03 AM Ticket #24 (Mechanische Schutzklasse) updated by Erik Welander
Revieweraccept changed
9:03 AM Ticket #24 (Mechanische Schutzklasse) updated by Erik Welander
Reviewerlist, Status changed
8:58 AM Ticket #23 (IP-Schutzklasse) updated by Erik Welander
Revieweraccept changed
8:58 AM Ticket #23 (IP-Schutzklasse) updated by Erik Welander
Reviewerlist, Status changed
8:54 AM Ticket #21 (Wiederherstellen der Bereitschaft nach Spannungsausfall) updated by Erik Welander
Revieweraccept changed
8:54 AM Ticket #21 (Wiederherstellen der Bereitschaft nach Spannungsausfall) updated by Erik Welander
Description, Reviewerlist, Status changed
8:51 AM Ticket #20 (Automatisches Wiedereinschalten nach Stromausfall) updated by Erik Welander
Revieweraccept changed
8:51 AM Ticket #20 (Automatisches Wiedereinschalten nach Stromausfall) updated by Erik Welander
Description, Reviewerlist, Reporter, Status changed
8:49 AM Ticket #19 (Absicherung gegen Stromausfall) updated by Erik Welander
Revieweraccept changed
8:49 AM Ticket #19 (Absicherung gegen Stromausfall) updated by Erik Welander
Description, Reviewerlist, Status changed
Die Aussage "mindestens 24-48h" bedeutet de Facto "mindestens 24h".
8:43 AM Ticket #18 (Externe Spannungsversorgung) updated by Erik Welander
Revieweraccept changed
8:43 AM Ticket #18 (Externe Spannungsversorgung) updated by Erik Welander
Description, Reviewerlist, Status changed
Die Maximalleistung von 15 W wurde in Besprechung mit Markus als …
8:34 AM Ticket #17 (Messbare Gase) updated by Erik Welander
Revieweraccept changed
8:34 AM Ticket #17 (Messbare Gase) updated by Erik Welander
Reviewerlist, Status changed
8:33 AM Ticket #16 (Maximaldurchfluss) updated by Erik Welander
Description, Reporter changed
8:31 AM Ticket #15 (Messbereich) updated by Erik Welander
Revieweraccept changed
8:31 AM Ticket #15 (Messbereich) updated by Erik Welander
Description, Reviewerlist, Status changed
8:29 AM Ticket #12 (PC über Netzwerk anschließen) updated by Erik Welander
Revieweraccept changed
8:29 AM Ticket #9 (Genauigkeit und Messbereich) updated by Erik Welander
Revieweraccept changed
8:29 AM Ticket #14 (Gastemperatur) updated by Erik Welander
Revieweraccept changed
8:29 AM Ticket #14 (Gastemperatur) updated by Erik Welander
Reviewerlist, Status changed
8:28 AM Ticket #13 (Genauigkeit nach OIML R 137-1 & 2 (2014)) updated by Erik Welander
Reviewerlist, Status changed
Im ursprünglichen Lastenheft von Michael Grexa wurde Klasse 0.5 ohne …
8:19 AM Ticket #12 (PC über Netzwerk anschließen) updated by Erik Welander
Reviewerlist, Status changed
7:05 AM Ticket #9 (Genauigkeit und Messbereich) updated by Erik Welander
Description, Reviewerlist, Reporter, Status changed
6:57 AM Ticket #8 (Betriebsvoraussetzungen) updated by Erik Welander
Revieweraccept changed
6:57 AM Ticket #8 (Betriebsvoraussetzungen) updated by Erik Welander
Summary, Description, Reviewerlist, Reporter, Status changed
6:45 AM Ticket #5 (Umwerter über Netzwerk anschließen) updated by Erik Welander
Revieweraccept changed
6:45 AM Ticket #5 (Umwerter über Netzwerk anschließen) updated by Erik Welander
Reviewerlist changed
6:44 AM Ticket #5 (Umwerter über Netzwerk anschließen) updated by Erik Welander
Description, Status changed

Aug 28, 2024:

3:23 PM Ticket #32 (Daten direkt auslesen) updated by Rotaru, Sebastian
Status changed
3:22 PM Ticket #3 (Umwerter remote anschließen) updated by Rotaru, Sebastian
Status changed
3:22 PM Ticket #2 (Zugriff Remote) updated by Rotaru, Sebastian
Status changed
3:21 PM Ticket #1 (Kommunikation) updated by Rotaru, Sebastian
Status changed
2:55 PM Ticket #126 (Figerprint nach ISO 17089) updated by Rotaru, Sebastian
Parent changed
2:51 PM Ticket #126 (Figerprint nach ISO 17089) created
Das Gerät soll einen Vergleichsreport nach ISO 17089 erstellen können. …
11:36 AM Ticket #89 (Signalqualität) updated by Erik Welander
Description changed
11:26 AM Ticket #32 (Daten direkt auslesen) updated by Erik Welander
Revieweraccept changed
11:25 AM Ticket #3 (Umwerter remote anschließen) updated by Erik Welander
Revieweraccept changed
11:25 AM Ticket #2 (Zugriff Remote) updated by Erik Welander
Revieweraccept changed
11:23 AM Ticket #3 (Umwerter remote anschließen) updated by Erik Welander
Reviewerlist, Status changed
11:22 AM Ticket #2 (Zugriff Remote) updated by Erik Welander
Description, Reviewerlist, Status changed
11:18 AM Ticket #32 (Daten direkt auslesen) updated by Erik Welander
Description, Reviewerlist, Reporter, Status changed
Display ist ein Muss für diese Art Zähler.
11:12 AM Ticket #89 (Signalqualität) updated by Jan-Hendrik Otto
11:05 AM Ticket #1 (Kommunikation) updated by Erik Welander
Revieweraccept changed
11:04 AM Ticket #1 (Kommunikation) updated by Erik Welander
Description, Reviewerlist, Status changed
9:53 AM Ticket #125 (Kosten für Steuerelektronik USA) created
Die Kosten für Steuerelektronik dürfen nicht 986 € übersteigen.
9:52 AM Ticket #124 (Kosten für Steuerelektronik ATEX / IECEx) created
Die Kosten für Steuerelektronik dürfen nicht 986 € übersteigen.
9:50 AM Ticket #117 (Kosten für Elektronikgehäuse USA) updated by Erik Welander
Description, Reporter changed
9:49 AM Ticket #116 (Kosten für Elektronikgehäuse ATEX / IECEx) updated by Erik Welander
Description, Reporter changed
9:45 AM Ticket #123 (Kosten für Display und Tastatur USA) created
Die Kosten für Display und Tastatur dürfen nicht 74 € übersteigen.
9:44 AM Ticket #122 (Kosten für Display und Tastatur ATEX / IECEx) created
Die Kosten für Display und Tastatur dürfen nicht 74 € übersteigen.
9:31 AM Ticket #121 (Kosten Rohrbaum USA) created
Die Kosten für den Rohrbaum inklusive Abdeckung wenn vorhanden dürfen …
9:30 AM Ticket #120 (Kosten Rohrbaum ATEX / IECEx) created
Die Kosten für den Rohrbaum inklusive Abdeckung wenn vorhanden dürfen …
9:22 AM Ticket #119 (Kosten für Sensoren USA) created
Die Kosten für die Sensoren (12 x) dürfen nicht 1957 € übersteigen.
9:21 AM Ticket #118 (Kosten für Sensoren ATEX / IECEx) created
Die Kosten für die Sensoren (12 x) dürfen nicht 1957 € übersteigen.
9:15 AM Ticket #117 (Kosten für Elektronikgehäuse USA) created
Die Kosten für alle Elektronikgehäuse, inklusive Klemmen und sonstige …
9:13 AM Ticket #116 (Kosten für Elektronikgehäuse ATEX / IECEx) created
Die Kosten für alle Elektronikgehäuse, inklusive Klemmen und sonstige …
9:07 AM Ticket #114 (Kosten Bauteile ATEX / IECEx) updated by Erik Welander
Description, Reporter changed
9:07 AM Ticket #115 (Kosten Bauteile USA) created
Die Gesamtkosten für die Bauteile für USA dürfen nicht 4645 € …
9:03 AM Ticket #114 (Kosten Bauteile ATEX / IECEx) created
Die Gesamtkosten für die Bauteile für ATEX / IECEx dürfen nicht 4035 € …
8:54 AM Ticket #113 (Bauteilenkosten) created
Metanforderung für Kosten für Bauteile.
8:27 AM Ticket #95 (Elektronikgehäuse) updated by Erik Welander
Depending on changed
Hinweis auf Anf. #55 entfernt, da nicht mehr aktiv.
8:26 AM Ticket #55 (Position des Displays) closed by Erik Welander
wontimplement: Aus Diskussion mit Sascha: wird nicht als Anforderung festgelegt.
8:22 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Description changed
Entscheidung im Meeting am 27.08.2024: Maximal 8 Pfade. Kommentar im …
8:18 AM Ticket #23 (IP-Schutzklasse) updated by Erik Welander
Description, Priority changed
Entscheidung im Meeting am 27.08.2024: IP-66
8:11 AM Ticket #66 (Sensorenanregung) updated by Erik Welander
Description changed
Anpassung an Ergebnis aus Meeting am 20.08.2024.
8:08 AM Ticket #110 (Anschluss für Temperatursensor) updated by Erik Welander
Im Meeting am 27.08.2024 wurde betont, dass ein Temperatursensor für …

Aug 27, 2024:

9:56 AM Ticket #112 (Drahtlose Kommunikation) updated by Erik Welander
Reporter changed
Entscheidung in Meeting von 28.05.2024.
9:51 AM Ticket #17 (Messbare Gase) updated by Erik Welander
Description changed
Grenze bis zu 30% Wasserstoff in Dokument …
9:41 AM Ticket #112 (Drahtlose Kommunikation) created
Das Gerät soll eine drahtlose Kommunikationsmöglichkeit, wie zum …
8:24 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Description changed
Kommentar "Class 0.5 Wiederholbarkeit 0.05 vielleicht nur möglich mit …
8:23 AM Ticket #111 (Wiederholbarkeit) updated by Erik Welander
Description changed
8:18 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Description changed
Kommentar aus Workshop "Release 1: Performance von GT400" wurde …
8:17 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Description changed
Diskussion aus Workshop "Wiederholbarkeit ist Bedingungsabhängig. …
8:17 AM Ticket #111 (Wiederholbarkeit) updated by Erik Welander
Description, Reporter changed
8:15 AM Ticket #111 (Wiederholbarkeit) created
Das Gerät mit 6 Pfaden, Class 1.0, soll mindestens eine …
7:52 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Description changed
Kommentar "Gleichrichter können Druckverlust verursachen." wurde in …
7:52 AM Ticket #71 (Gleichrichter) updated by Erik Welander
Description changed
7:46 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Description changed
7:42 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Description changed
Aus Workshop "Die Hauptfrage: Gleichrichter erlaubt oder nicht? …
7:38 AM Ticket #71 (Gleichrichter) updated by Erik Welander
Description, Reporter changed
7:29 AM Ticket #13 (Genauigkeit nach OIML R 137-1 & 2 (2014)) updated by Erik Welander
Description changed
7:28 AM Ticket #13 (Genauigkeit nach OIML R 137-1 & 2 (2014)) updated by Erik Welander
Description changed
7:24 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Description changed
7:14 AM Ticket #3 (Umwerter remote anschließen) updated by Erik Welander
Description changed
7:05 AM Ticket #48 (Pfadlanzahl) updated by Erik Welander
Description changed

Aug 26, 2024:

7:06 AM Ticket #26 (Zugriff Lokal) updated by Erik Welander
Description changed
6:37 AM Ticket #110 (Anschluss für Temperatursensor) updated by Erik Welander
Der Temperatursensor soll für die Reynoldskorrektur und Kalibrierung …
6:32 AM Ticket #25 (Elektromagnetische Schutzklasse) updated by Erik Welander
Description, Reporter changed

Aug 22, 2024:

10:26 AM Ticket #108 (Optionale Signaldurchfürhung) updated by Erik Welander
Description, Depending on, Reporter changed
10:21 AM Ticket #73 (Durchflusskorrekturen) updated by Erik Welander
Depending on changed
10:20 AM Ticket #73 (Durchflusskorrekturen) updated by Erik Welander
Depending on, Reporter changed
10:20 AM Ticket #110 (Anschluss für Temperatursensor) updated by Erik Welander
Description, Reporter changed
10:07 AM Ticket #105 (Integrierte Temperatur und Drucksensoren) closed by Erik Welander
wontimplement: Wird durch externe Signalführung umgesetzt.
10:02 AM Ticket #110 (Anschluss für Temperatursensor) created
Das Gerät muss einen Anschluss (Thermowell) für Temperatursensor …
8:53 AM Ticket #19 (Absicherung gegen Stromausfall) updated by Erik Welander
Description changed
Die Frage mit welchen Einschränkungen das Gerät im Batteriebetrieb …
8:52 AM Ticket #109 (Funktionsumfang im Backupbetrieb) created
Das Gerät soll im Backupbetrieb den gesamten Funktionsumfang weiterhin …
8:43 AM Ticket #104 (Anschluss für PT100) updated by Erik Welander
Description, Parent, Class, Reporter changed
Update aus RSM420_AnschlussMatrix_Rev1_20240304.xlsx, Stand 22.08.2024.
8:36 AM Ticket #55 (Position des Displays) updated by Erik Welander
Description changed
Die Entscheidung ein externes Display zu verwenden, wurde im Team …
8:34 AM Ticket #100 (Auslegung Backup Stromversorgung) updated by Erik Welander
Description, Priority changed
Die Entscheidung Akkumulatoren zu verwenden, wurde als bevorzugter Weg …
7:30 AM Ticket #11 (Frequenzausgänge) updated by Erik Welander
Description changed
Dies entspricht Dokument RSM420-AnschlussMatrix_Rev1_20240304.xlsx, …

Aug 20, 2024:

3:16 PM Ticket #101 (Minimaler Innenraum) updated by Erik Welander
Description, Class, Reporter changed
Dies entspricht dem Innenraum von Cortem CCA-02E, in Meeting von …
2:39 PM Ticket #66 (Sensorenanregung) updated by koerner
Replying to (keine): > Die Anregefrequenz für die Sensoren …
2:35 PM Ticket #66 (Sensorenanregung) updated by koerner
Auf Grund von CO2 und H2 muss die Anregefrequenz im Bereich 50 - 500 …
Note: See TracTimeline for information about the timeline view.